Probes and Biopsy 12-37
(2) Turn on the knob of fixing guiding block and remove the guiding block from the hole
of installing guiding block.
(3) Turn on the knob of fixing needle-guided bracket and remove the needle-guided
bracket.
NGB-010
(1) Remove the guiding block slightly along the direction of the needle’s tail, and
separate the residual part of the needle-guide bracket and the transducer from the
needle.
(2) Remove the support of needle-guided bracket from the transducer.
NGB-016
1. Screw the nut of the guiding block and remove the guiding block slightly along the
direction of the needle’s tail.
2. Separate the residual part of the needle-guide bracket and the transducer from the
needle.
3. Screw the pinch nut of the bracket, and remove the needle-guided bracket from the
transducer.
4. Separate the transducer and the needle-guided bracket.
12.2.9 Clean and Sterilize the Needle-guided Bracket
Cleaning
1. Wear sterile gloves to prevent infection.
2. Or, wash with clean water or soapy water to remove all the foreign matters, or, wipe the
bracket with a soft ethyl carbamate sponge.
3. Wipe off the water on the needle-guided bracket using sterile cloth or gauze after
washing it.
Please follow the instructions in the manual for cleaning.
Sterilization
1. Wear sterile gloves to prevent infection.