Acoustic Output Reporting Table (60601-2-37) C-23
C.4 65C15EA
Transducer Model: 65C15EA
Imaging Mode: M-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.78
0.05
0.23
0.22
Index component value
0.05
0.03
0.13
0.23
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
1.71
P (mW)
2.33
3.37
7.62
P
1
(mW)
2.27
3.37
z
s
(cm)
1.44
z
b
(cm)
1.08
z
MI
(cm)
1.65
z
pii
,α
(cm)
1.65
f
awf
(MHz)
4.86
4.67
4.76
4.66
Other
Informatio
n
prr (Hz)
2,999.0
0
srr (Hz)
/
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
132.42
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
169.14
I
spta
at z
pii
or z
sii
(mW/cm
2
)
298.60
p
r
at z
pii
(MPa)
2.06
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
12.9cm
12.9cm
12.9cm
12.9cm
Focus position
1.5cm
5.0cm
2.5cm
5.0cm
Working Frequency
H9.0
H9.0
H9.0
H9.0
PRF
2999
2999
2999
2999

Table of Contents