C-24 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 65C15EA
Imaging Mode: B-mode/Smart3D/Tissue Harmonic Imaging
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.77
0.11
0.21
0.21
Index component value
0.11
0.11
0.21
0.11
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
1.69
P (mW)
12.00
12.00
12.00
P
1
(mW)
4.69
4.69
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
1.65
z
pii
,α
(cm)
1.65
f
awf
(MHz)
4.84
4.81
4.81
4.72
Other
Informatio
n
prr (Hz)
5,096.0
0
srr (Hz)
13.00
n
pps
1.33
I
pa,α
at z
pii
,α
(W/cm
2
)
126.11
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
2.04
I
spta
at z
pii
or z
sii
(mW/cm
2
)
3.72
p
r
at z
pii
(MPa)
2.19
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
12.9cm
12.9cm
12.9cm
12.9cm
Focus position
1.5cm
10.0cm
10.0cm
5.0cm
Working Frequency
H9.0
H9.0
H9.0
H9.0
PRF
5096
5096
5096
5096

Table of Contents