EasyManua.ls Logo

Mindray DP-50 Vet - Page 273

Mindray DP-50 Vet
343 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Acoustic Output Reporting Table (60601-2-37) C-25
Transducer Model: 65C15EA
Imaging Mode: B+M-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.92
0.21
0.36
0.54
Index component value
0.21
0.19
0.25
0.36
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
2.03
P (mW)
20.32
19.64
20.35
P
1
(mW)
9.30
11.11
z
s
(cm)
1.44
z
b
(cm)
1.75
z
MI
(cm)
1.56
z
pii
,α
(cm)
1.56
f
awf
(MHz)
4.83
4.70
4.75
4.82
Other
Informatio
n
prr (Hz)
750.00
srr (Hz)
10.00
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
193.95
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
59.84
I
spta
at z
pii
or z
sii
(mW/cm
2
)
103.13
p
r
at z
pii
(MPa)
2.25
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
12.9cm
12.9cm
6.5cm
6.5cm
B/M Focus position
1.5cm
5.0cm
1.0cm
1.0cm
B/M Working Frequency
H9.0
H9.0
H9.0
H9.0
B PRF
4250
4250
6000
6000
M PRF
750
750
3000
3000

Table of Contents

Related product manuals