C-26 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 65C15EA
Imaging Mode: PW-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
1.16
0.04
0.13
0.07
Index component value
0.04
0.02
0.09
0.13
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
2.60
P (mW)
1.56
2.04
1.39
P
1
(mW)
1.52
2.04
z
s
(cm)
1.44
z
b
(cm)
1.56
z
MI
(cm)
1.36
z
pii
,α
(cm)
1.36
f
awf
(MHz)
5.02
5.01
5.03
5.03
Other
Informatio
n
prr (Hz)
699.00
srr (Hz)
/
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
342.45
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
180.40
I
spta
at z
pii
or z
sii
(mW/cm
2
)
308.55
p
r
at z
pii
(MPa)
2.91
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
6.5cm
12.9cm
6.5cm
6.5cm
SV Position
1.0cm
10.0cm
1.0cm
0.5cm
Working Frequency
5
5
5
5
PRF
699
6300
23995
23995
SV
0.5mm
0.5mm
0.5mm
0.5mm

Table of Contents