C-32 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 65EB10EA_T
Imaging Mode: B+M-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.86
0.05
0.10
0.11
Index component value
0.05
0.05
0.07
0.10
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
2.11
P (mW)
4.46
6.45
4.45
P
1
(mW)
2.24
3.12
z
s
(cm)
0.79
z
b
(cm)
0.79
z
MI
(cm)
1.65
z
pii
,α
(cm)
1.65
f
awf
(MHz)
6.03
5.03
4.99
5.00
Other
Informatio
n
prr (Hz)
1,000.0
0
srr (Hz)
32.00
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
169.87
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
39.05
I
spta
at z
pii
or z
sii
(mW/cm
2
)
81.87
p
r
at z
pii
(MPa)
2.75
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
12.9cm
6.5cm
6.5cm
6.5cm
B/M Focus position
2.0cm
0.5cm
0.5cm
0.5cm
B/M Working Frequency
6.5
5
5
5
B PRF
4000
8000
8000
8000
M PRF
1000
1000
1000
1000

Table of Contents