C-70 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 75L53EA
Imaging Mode: Color+B-Mode / Power+B-Mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.94
0.13
0.13
0.35
Index component value
0.13
0.13
0.13
0.13
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
2.24
P (mW)
24.00
24.00
26.00
P
1
(mW)
4.52
4.52
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
1.17
z
pii
,α
(cm)
1.17
f
awf
(MHz)
5.61
6.48
6.49
5.10
Other
Information
prr (Hz)
6,341.00
srr (Hz)
23.00
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
267.48
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
11.59
I
spta
at z
pii
or z
sii
(mW/cm
2
)
18.25
p
r
at z
pii
(MPa)
2.80
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
8.3cm
8.3cm
8.3cm
8.3cm
B Focus Position
1.5cm
5.0cm
5.0cm
5.0cm
Color Sampling Gate
Position
1.5cm
5.0cm
5.0cm
5.0cm
B Working Frequency
7.5
7.5
7.5
5
C Working Frequency
6.6
6.6
6.6
5.7
B PRF
3800
3039
3039
2807
Color PRF
6341
5333
5333
4926