C-2
Item Name O.M General User Defaults
ISO * * -80 ms
J 48 ms
ST J + 60 ms
X represents I, II, III, aVR, aVL, aVF, V.
Arrh. Analysis
Item Name
O.M
General User Defaults
C M
Arrhythmia Threshold Settings
PVCs High * * 10
Tachy * * 120
Brady * * 50
Asys. Delay * * 5
Vtac Rate * * 130
Vtac PVCs * * 6
Multif. PVC’s Window * * 15
Extreme Tachy * * 160
Extreme Brady * * 35
Vbrd Rate * * 40
Vbrd PVCs * * 5
Pause Time * * 2
Arrhythmia Alarm Settings
PVCs/min Alarm * * Off
R on T Alarm * * Off
Nonsus. Vtac Alarm * * Off
Vent. Rhythm Alarm * * Off
Bigeminy Alarm * * Off
Trigeminy Alarm * * Off
Asystole Alarm * * On
VFib/VTac Alarm * * On
Vtac Alarm * * On
Vent. Brady Alarm * * On
Extreme Tachy Alarm * * On
Extreme Brady Alarm * * On
X Alarm * * Off
Asystole Alm Lev * * High
VFib/VTac Alm Lev * * High
Vtac Alm Lev * * High
Vent. Brady Alm Lev * * High
Extreme Tachy Alm
Lev
* * High
Extreme Brady Alm
Lev
* * High
VT>2 Alm Lev * * Low