D-6
Measurement Alarm Message L I Cause and Solution
Recorder Unavailable L A
Recorder Vlt High L C An error occurred to the system power supply. Restart the
monitor.
Recorder Vlt Low L C
Recorder Head Hot L C The recorder has been working for too long time. Stop the
recording and resume the recording till the recorder’s
printhead cools down.
Rec Paper Wrong
Pos.
L A Re-load the recorder paper.
System System Watchdog Err H C An error occurred to the system. Restart the monitor.
System Software Err H C
System CMOS Full H C
System CMOS Err H C
System FPGA Err H C
System Err N H C
N is within 2 to 12.