7-7
Module
Parameter Low alarm limit
High alarm limit
Auto alarm limits range
(Cuff Type: Big)
NIBP-D Dia×0.68 + 6mmHg
Dia×0.86 + 32mmHg
2
5 to 210
NIBP-M Mean×0.68 + 8mmHg
Mean×0.86 + 35mmHg
30
to 230
Temp
T1 (T1 0.5) °C
(T1 + 0.5)
°C 1 to 49 °C
T2 (T2 0.5) °C
(T2 + 0.5)
°C 1 to 49 °C
TD
Same as the default alarm
limit
Same as the default alarm
limit
Same as the measurement range
IBP
ART/
Ao/
UAP/
BAP/
FAP/
LV/
P1
-P4
(Arterial
pressure)
IBP-S
(SYS15) or 45mmHg
(whichever is greater)
(
SYS + 15) or 105mmHg
(
whichever is smaller)
35 to 115
IBP-D
(Dia15) or 20mmHg
(whichever is greater)
(
Dia + 15) or 80mmHg
(whichever is smaller)
20 to 90
IBP-M
(Mean15) or 35mmHg
(whichever is greater)
(
Mean + 15) or 95mmHg
(whichever is smaller)
25 to 105
PA
IBP-S SYS × 0.75
SYS
× 1.25
3 to 120mmHg IBP-D Dia × 0.75
Dia
× 1.25
IBP-M Mean × 0.75
Mean
× 1.25
IBP:
CVP/
ICP/
LAP/
RAP/
UVP/
P1
-P4
(Venous
pressure)
IBP-M Mean × 0.75
Mean
× 1.25 3 to 40mmHg
CO
2
EtCO
2
0 to 32mmHg:
remains the same
0
to 32mmHg:
remains the same
Same as the measurement range
32 to 35mmHg:
29mmHg
32
to 35mmHg:
41mmHg
35 to 45mmHg:
(etCO
2
-6) mmHg
35
to 45mmHg:
(etCO
2
+6) mmHg
45 to 48mmHg:39 mmHg
45
to 48mmHg:51 mmHg
>48mmHg: remains the
same
>48mmHg:
remains the same
FiCO
2
N/A
Same as the default alarm limit
Same as the measurement range
awRR
(awRR10) or 30 rpm
(whichever is greater)
(
awRR+25) or 85 rpm
(whichever is smaller)
10 to 90