A-20
Alarm limit Range Step
TB High
(low limit + 1) to 43 °C
(low limit + 1.8) to 109.4 °F
TB Low
23 to (high limit - 1) °C
73.4 to (high limit - 1.8) °F
0.1 °C
0.1 °F
A.5.9 CO
2
Standard Meet standard of ISO 21647
Measurement mode Sidestream, microstream, mainstream
Technique Infrared absorption
Sidestream CO
2
Module
CO
2
Measurement range 0 to 99 mmHg
Accuracy*
0 to 40 mmHg:
41 to 76 mmHg:
77 to 99 mmHg:
±2 mmHg
±5% of the reading
±10% of the reading
Accuracy drift Meet the requirement for measurement accuracy within 6 hours
Resolution 1 mmHg
Sample flowrate 70 ml/min, 100 ml/min 120 ml/min, 150 ml/min
Sample flowrate tolerance ±15% or ±15 ml/min, whichever is greater.
Warm-up time 1 min
Response time
Measured with a neonatal watertrap and a 2.5-meter neonatal
sampling line:
<3 s @ 150 ml/min
<3 s @ 120 ml/min
<3.5 s @ 100 ml/min
<4 s @ 70 ml/min
Measured with a neonatal watertrap and a 2.5-meter adult
sampling line:
<4.5 s @ 150 ml/min
<5.5 s @ 120 ml/min
<5.5 s @ 100 ml/min
<7 s @ 70 ml/min
Gas sampling delay time
Measured with a neonatal watertrap and a 2.5-meter neonatal
sampling line:
<2.5 s @ 150 ml/min
<3 s @ 120 ml/min
<3 s @ 100 ml/min
<3.5 s @ 70 ml/min