A-23
Mainstream CO
2
Module
CO
2
Measurement range 0 to 150 mmHg
Accuracy
0 to 40 mmHg:
41 to 70 mmHg:
71 to 100 mmHg:
101 to 150 mmHg:
±2 mmHg
±5% of the reading
±8% of the reading
±10% of the reading
Accuracy drift Meet the requirement for measurement accuracy within 6 hours
Resolution
0 to 69 mmHg:
70 to 150 mmHg:
1 mmHg
0.25 mmHg
Rise time <60 ms
awRR measurement range 0 to 150 rpm
awRR measurement accuracy ±1 rpm
Alarm limit Range Step
EtCO2 High (low limit + 2) to 150 mmHg
EtCO2 Low 0 to (high limit - 2)mmHg
FiCO2 High 0 to 150 mmHg
1 mmHg
awRR High (low limit + 2) to 100 rpm
awRR Low 6 to (high limit - 2) rpm
1 rpm
A.5.10 AG
Standards Meet standard of ISO 21647 and YY0601
Technique Infrared absorption
Warm-up time
Iso accuracy mode:
Full accuracy mode:
45 s
10 min
Sample flow rate
Adult, pediatric:
Neonate:
Accuracy:
120, 150, 200 ml/min
70, 90, 120 ml/min
±10 ml/min or ±10%, whichever is greater
Measurement range
CO
2
:
O
2,
N
2
O:
Des:
Sev:
Enf, Iso, Hal:
awRR:
0 to 10%
0 to 100%
0 to 18%
0 to 8%
0 to 5%
2 to 100 rpm
Resolution
CO
2
:
awRR:
1 mmHg
1 rpm