The following picture takes O2+N2O+Air configuration as an example to illustrate how
pipeline gas supplies are outputted. O2 is divided into two pathways (into three pathways if
auxiliary O2 supply is configured: system switch 23, O2 flush valve 19 and auxiliary O2
supply 46 respectively). One pathway of O2 flows into system switch 23 and the other into
O2 flush valve 19. N2O flows into O2-N2O cut-off valve 25 and Air into system switch 23.
N2O pipeline supply inlet
assembly
O2 pipeline
supply inlet
assembly
Air pipeline
supply inlet
assembl
y
N2O tube
O2 tube
Air tube
25. O2-N2O
cut-off valve
23.System
switch
19. O2 flush
valve