7-32
For FPM as a calibration device, prepare the following plugs and sampling ports. Connect the
plugs, sampling ports and FPM inspiratory flow sensor ports as shown in the following figure.
Note that the tube length must not exceed 0.6m.
6. Connect the power supply to the calibration device and zero the calibration device
manually.
7. Configure the calibration device.
For details, see step 5 in section 4.3.2.1.
8. Press [Next]. The menu as shown in the following figure is displayed: