3.4Warmup time
To deliver exact results, the weighing platform must warm up for at least 30 minutes after
initial connection to AC power or after a relatively long power outage. Only after this time
will the weighing platform have reached the required operating temperature.
Preparing the Weighing Platform for Verication for Use in Legal Metrology in the
EU:*
* including the Signatories of the Agreement on the European Economic Area
Allow the weighing platform to warm up for at least 24 hours after initial connection to
the power supply.
3.5Menu access switch
Service switch
Service switch right: Conguration enabled
Service switch left: Conguration blocked (verication access switch)
Menu Access Switch Prior to Verication:
-All settings in the operating menu are accessible (once the weighing platform has
been veried for use in legal metrology, menu items relevant to legal metrology are
blocked)
-Select conguration settings from corresponding specications group (conguration
cannot be changed after verication)
-Block the ‘write’ function for the xBPI interface
Function of the Menu Access Switch for Verication as a Legal Measuring Instrument:
-When the menu access switch (9) is open, weighing ranges can be congured and
calibration/adjustment can be performed, prior to verication.
Function of the Menu Access Switch after verication:
-Move the menu access switch to the left. Conguration settings relevant to use of the
weighing platform in legal metrology are no longer accessible.
3.6Operating limits
Maximum Overload Capacity
Because Minebea Intec weighing platforms are constructed to withstand rough treatment
in industrial environments, the platform will not be damaged if occasionally loaded
beyond the maximum weighing capacity.
3 Setup instructionsISEDE‑16‑34‑64‑H(V1) | ISEDE‑16‑34‑HCE | ISEDE‑64‑S(V1) | ISEDE‑64‑SCE
EN-11Minebea Intec