7Error messages
Error messages are output via the interface or control unit.
Display
Info/ERR
CauseRemedy
01Data output not compatible with output format.Set output format correctly.
02Calibration/adjustment condition not met, e.g.
not tared or the scale is unloaded.
Calibrate only when zero is displayed.
03Calibration/adjustment could not be completed
within a certain time.
Make sure there is no draft and repeat the cali-
bration/adjustment process.
06Built-in CAL weight defective.Contact Minebea Intec Customer Service.
07The last function activated is not allowed for sca-
les veried for use in legal metrology.
Contact Minebea Intec Customer Service for in-
formation on changing settings.
08The load on the scale is too heavy to zero the
readout.
Check whether the "initial zero setting range"
was set properly in your conguration.
09Taring is not possible when the gross value is ≤ 0.Zero the scale.
10Tare key is blocked when there is data in the pre-
set tare memory.
The preset tare memory must rst be deleted to
restore the tare function.
50Overow/underow of TC converterContact Minebea Intec Customer Service.
53Failure of TC converterContact Minebea Intec Customer Service.
54/LModulation of weighing converter too lowPosition the weighing pan.
55/HModulation of weighing converter too highUnload the scale.
ISEDE‑16‑34‑64‑H(V1) | ISEDE‑16‑34‑HCE | ISEDE‑64‑S(V1) | ISEDE‑64‑SCE7 Error messages