Minebea Intec PR 5230 - Limit Parameter; Digital I;O Parameters; Analog Output Parameter

Minebea Intec PR 5230
303 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Limit 13 on
Enter 0 – Max (maximum load); take unit from
calibration.
Action
Action, selection: no action, set marker 13, clr
(clear) marker 13
Condition Condition, selection: see Chapter 7.15.8.
Limit 13 o
Enter 0 – Max (maximum load); take unit from
calibration.
Action
Action, selection: no action, set marker 13, clr
(clear) marker 13
Condition Condition, selection: see Chapter 7.15.8.
Output 1…3 Congure outputs, selection: see
Chapter 7.15.10.
Input 1…3 on Congure inputs
Action Action, selection: see Chapter 7.15.9.1.
Condition Condition, selection: see Chapter 7.15.9.2.
Input 1…3 o Congure inputs
Action Action, selection: see Chapter 7.15.9.1.
Condition Condition, selection: see Chapter 7.15.9.2.
Analog mode Analog output parameter: Gross D08 = gross,
Net if tared D09 = Net if tared, Selected D11 =
value on the display, Transparent D30 =
Transparent, no output = analog output
unused, see Chapter 7.15.12.
Analog range
Output range: 020 mA, <420 mA>
Output on error Output on error: 0 mA = set to 0 mA, <4 mA> =
set to 4 mA, 20 mA = set to 20 mA, hold = last
output value remains unchanged
Output if <0 Output if <0: set 0 mA = 0 mA, set <4 mA> =
4 mA, set 20 mA = 20 mA, linear = goes below
4 mA down to the limit (with 420 mA)
Output if >Max
7.11.9 Limit parameter
Note:
This menu item is only available if under -[Operating parameter]- [Application]
"Standard" has been selected.
7.11.10 Digital I/O parameters
Note:
This menu item is only available if under -[Operating parameter]- [Application]
"Standard" has been selected.
7.11.11 Analog output parameter
7 Commissioning Transmitter in eld housing PR 5230
EN-123 Minebea Intec