EasyManua.ls Logo

Minebea Intec PR 5230 - Calibrating with Weight

Minebea Intec PR 5230
303 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
To use the empty scale/hopper as dead load (normal case):
Note:
If the mV/V value of the dead load was calculated, or if it is known from the previous
calibration, the value can be overwritten by pressing [by mV/V].
Error messages, calibration see Chapter 16.5.
7.12.7 Calibrating with weight
Note:
If a linearization was carried out (see Chapter 7.12.11), after selection of the line [Max at]
the following tip is displayed:
Cannot be changed here
while linearization is active.
Changes cannot be made while linearization is switched on.
Only deleting of the linearization points deactivates the linearization mode!
1. Clear the scale/hopper.
2. Press the [by load] softkey.
3. Conrm entries with or .
The verication is displayed by "Setting dead load…".
7 Commissioning Transmitter in eld housing PR 5230
EN-131 Minebea Intec

Table of Contents

Related product manuals