The weight unit for the CAL weight may dier from the unit in the device. Conversion
is automatic.
Note:
Error messages, calibration see Chapter 16.5.
7.12.8 Calibrating with mV/V
The scale can be calibrated without weights. During input of the load cell mV/V value, the
acceleration of gravity at the place of installation can be taken into account.
The STAR load cell data is based on the acceleration of gravity in Hamburg, Germany:
9.81379 m/s
2
.
1. Press the [by load] softkey.
2. Place the CAL weight on the scale.
3. Enter the weight of the CAL weight.
4. Conrm the entries.
5. Press to select the weight unit.
6. Conrm entries with or .
Weight value, weight unit and measuring signal in mV/V corresponding to this
value are displayed in the line [Calibrated at].
The verication is displayed by "Setting Span by load…".