To view the range which is already utilized by zero tracking or zero setting, in [Calibration]
press the key; this also activates 10-fold increased resolution of the weight value.
Note:
The scale must not be loaded!
If the full zero-setting range is already being utilized, you can still correct the dead load
(overwrite protection must be deactivated, see Chapter 7.1.3.1) without aecting other
calibration data/parameters. To do this select -[Weighing points]- [Calib]- [Modify]
and determine the dead load with [Dead load at] using the [by load] option (see
Chapter 7.12.6).
Note:
If a linearization was carried out (see Chapter 7.12.11), the following note appears on the
display after the [Dead load at] line is selected:
Cannot be changed here
while linearization is active.
Changes cannot be made while linearization is switched on.
Only deleting of the linearization points deactivates the linearization mode!
7.12.11 Linearization
The measurement range for a straight can be optimized by setting the linearization
points.
Requirements:
Calibration of Max and dead load was done.
Transmitter in eld housing PR 5230 7 Commissioning
Minebea Intec EN-136