- In the menu [Weighing point A] - [xBPI-Scale] - [Setup] at [Conguration] -
[Weighing parameters] - [Conrming adjust.] was set to "manual".
- The communication between the device and platform is active.
Procedure:
7.14 Calibrating digital load cells of type "Pendeo"
7.14.1 General information
The digital load cells have been calibrated at the factory based on the acceleration of
gravity at Hamburg (9.81379 m/s
2
). The calibration data in the load cells are invariable.
1. Select -[Weighing point]- [xBPI-Scale] and conrm.
2. Press the [Setup] softkey.
The parameters of the xBPI-scale are read into the device.
3. Select [Calibration]- [Linearity: Default] using the cursor and conrm.
The rst linearization point to be calibrated is displayed.
4. Place the displayed weight on the scale.
The deviation is displayed in the last line with increased resolution (10-fold).
5. Press the [Accept] softkey.
The second linearization point to be calibrated is displayed.
6. Place the displayed weight on the scale.
The deviation is displayed in the last line with increased resolution (10-fold).
7. Press the [Accept] softkey.
The third linearization point to be calibrated is displayed.
8. Place the displayed weight on the scale.
The deviation is displayed in the last line with increased resolution (10-fold).
9. Press the [Accept] softkey.
The last linearization point to be calibrated is displayed.
10. Place the displayed weight on the scale.
The deviation is displayed in the last line with increased resolution (10-fold).
11. Press the [Accept] softkey.
12. Press the softkey to exit the menu and to save the settings.
Transmitter in eld housing PR 5230 7 Commissioning