The rst line cannot be changed.
Possible settings, see Chapter 3.4.3.3.
7.15.8 Conguring limit values
Note:
This menu item is only available if under -[Operating parameter]- [Application]
"Standard" has been selected.
The parameters for limits are set under this menu item.
Each limit consists of a switch-on and a switch-o point for denition of a hysteresis. The
3 pairs of values must be entered according to the same principle. The limit values always
refer to the gross weight. For the SPM addresses for the limits, see Chapter 13.4.
NOTICE
The limit values of an xBPI weighing point are scale-specic.
The scale must be active when entering the limit values.
The scale and the unit must not be changed after conguration.
The following settings are required: [Weighingpoint/xBPI-Scale]- [Setup]-
[Conguration]- [Application settings]- [Number of units] "1 Weight"
Dene the parameters for limits under -[Limit parameter].
7 Commissioning Transmitter in eld housing PR 5230
EN-195 Minebea Intec