Parameter table
Menu item Selection Description
[no output] Analog output is unused.
[Gross D08] Output of the gross weight.
[Net if tared D09] Output of the net weight, if tared; otherwise gross
weight
[Selected D11] Output of the gross/net value on the display, de-
pendent on SPM bit X72.
[Analog mode]
[Transparent D30] Output of the value in D30
[020 mA] Output of 020 mA.
[Analog range]
[420 mA] Output of 420 mA.
[0 mA] Set output to 0 mA.
[4 mA] Set output to 4 mA.
[20 mA] Set output to 20 mA.
[Output on error]
[hold] The last output value is held.
[0 mA] Set output to 0 mA.
[4 mA] Set output to 4 mA.
[20 mA] Set output to 20 mA.
[Output if < 0]
[linear] The output drops below 4 mA down to the limit (at
420 mA).
[0 mA] Set output to 0 mA.
[4 mA] Set output to 4 mA.
[20 mA] Set output to 20 mA.
[Output if > Max]
[linear] Output increases above 20 mA until the limit is re-
ached.
[Weight at 0/4 mA] Weight value for 0/4 mA output.
[Weight at 20 mA] Weight value for 20 mA output.
Quit the menu and save parameters by pressing .
Note:
Adapting the analog output, see Chapter 9.1.2.1.
7.16 System information
This menu displays system information.
You can also check inputs and outputs, see Chapter 9.1.2.2.