16 Error messages
16.1 Error messages measuring circuit
The internal weighing electronics can generate error messages, which are shown on the
weight display.
Display VNC text Error and possible cause Remote dis-
play
E:Arith Arith. error Internal arithmetic overow:
- Faulty calibration values.
Error 1
E:Overl Overload The measuring signal is higher than Max + (x d):
- Wrong setting.
- Too much weight on the scale.
Error 2
E:NoSig Ext. meas. device error Measuring input open:
- The measuring signal is higher than the per-
missible range of 36 mV.
- Measuring cable is interrupted (cable break
detection).
- Other hardware defect.
Error 3
E:Under Value exceeds display The weight value is not displayed:
- Too many digits have been set.
Error 4
E:Sense * No sense voltage No sense voltage:
- Load cells not connected.
- Sense line or supply line is interrupted.
- Wrong polarity or sense voltage is low.
Error 6
E:Invers Negative input Negative measuring signal:
- Wrong polarity of load cell signal.
- Wrong polarity of supply voltage.
Error 7
E:HardE No values from scale Internal weighing point:
The measuring signal is higher than the permis-
sible range of 36 mV. Cannot read weight values
from ADC (analog-digital converter).
- Error in weighing electronics board.
- Defective load cell.
- Cable break.
Error 9
* with Ex applications strictly see Chapter 16.7.
16.2 Weight error status
Transmitter in eld housing PR 5230 16 Error messages
Minebea Intec EN-286