4.5.3 Connecting a load cell with a 6-wire cable
gn
gy
rd
wh
bk
bu
+
M
M
+
V
+
S
S
V
+ Meas.
– Meas.
+ Supply
+ Sense
– Sense
– Supply
Ground
GNDA
insert screen in the sleeve
equipontial bonding conductor
Terminal Port/color code Description
M+ + Meas./gn + Measuring voltage (load cell output)
M- - Meas./gy - Measuring voltage (load cell output)
S+ + Sense/wh + Sense voltage
S- - Sense/bk - Sense voltage
V+ + Supply/rd + Supply voltage
V- - Supply/bu - Supply voltage
GNDA Ground analog Screen (ground)