6. Operating | 6.2 - Configuration and adjustment via the keypad | 104
Saving and exiting
14. Press "Exit".
SAVE appears on the display.
15. Press "OK".
yES appears on the display.
16. Press "OK".
SAVE ... appears on the display while saving is in progress.
17. Press "Exit" to exit the setup menu.
18. If no further adjustment data or parameters are to be set, the CAL switch
must now be closed, see ChapterCAL switch.
Note:
If the mV/V value of the dead load was calculated, or if it is known from
the previous adjustment, the value can be overwritten by pressing byMV-V
(mV/V input).