6. Operating | 6.2 - Configuration and adjustment via the keypad | 125
1. Press and hold for 2–3 seconds.
SEtuP appears on the display.
2. Press "OK".
Cd 000 appears on the display (Cd flashes).
3. Press (3×) to select do 040.
4. Press "OK".
outP. 1 appears on the display.
5. Press "OK".
b00=0 appears on the display.
6. Press / to select the function for output 1, see selection list.
7. Perform the same steps for output 2 and output 3 in the menus do 041
and do 042.
8. Press "Exit".
9. Save the changes.
Example: Overload function
SPM bit [X34 = 1]
Function and output are active (e.g. if overload is reached, a bulb lights
up).
SPM bit [X34 = 0]
Function is active and output is inactive (e.g. if overload is reached, a bulb
goes out).