6. Operating | 6.2 - Configuration and adjustment via the keypad | 127
Parameter Selection Description
[linear] In the event of a weight >Max, the analog out
put is retained. This is only possible if the out
put value for Max is <20 mA.
[0mA] In the event of a weight >Max, the output is set
to 0mA.
[4mA] In the event of a weight >Max, the output is set
to 4mA.
[On > Max]
[20mA] In the event of a weight >Max, the output is set
to 20mA.
[Analog value]
Input: fixed analog val
ue
Only possible if [Transparent] mode is selected.
[Weight value at
0/4mA]
Input: Weight value Weight value at which the analog output
should display 0mA.
or
Weight value at which the analog output
should display 4mA (for the analog range 4…
20mA).
[Weight value at
20mA]
Input: Weight value Weight value at which the analog output
should display 20mA.
[Adjust for 4mA]
Input: Weight value Value at 4mA. Adjusting the analog output, see
Chapter Adapting the analog output.
[Adjust for 20mA]
Input: Weight value Value at 20mA. Adjusting the analog output,
see Chapter Adapting the analog output.
1. Press and hold for 2–3 seconds.
SEtuP appears on the display.
2. Press "OK".
Cd 000 appears on the display (Cd flashes).
3. Press (5×) to select Ao 050.
4. Press "OK".
AnA.Mod appears on the display.
5. Press "OK".

Table of Contents