6. Operating | 6.2 - Configuration and adjustment via the keypad | 127
Parameter Selection Description
[linear] In the event of a weight >Max, the analog out
put is retained. This is only possible if the out
put value for Max is <20 mA.
[0mA] In the event of a weight >Max, the output is set
to 0mA.
[4mA] In the event of a weight >Max, the output is set
to 4mA.
[On > Max]
[20mA] In the event of a weight >Max, the output is set
to 20mA.
[Analog value]
Input: fixed analog val
ue
Only possible if [Transparent] mode is selected.
[Weight value at
0/4mA]
Input: Weight value Weight value at which the analog output
should display 0mA.
or
Weight value at which the analog output
should display 4mA (for the analog range 4…
20mA).
[Weight value at
20mA]
Input: Weight value Weight value at which the analog output
should display 20mA.
[Adjust for 4mA]
Input: Weight value Value at 4mA. Adjusting the analog output, see
Chapter Adapting the analog output.
[Adjust for 20mA]
Input: Weight value Value at 20mA. Adjusting the analog output,
see Chapter Adapting the analog output.
1. Press and hold for 2–3 seconds.
SEtuP appears on the display.
2. Press "OK".
Cd 000 appears on the display (Cd flashes).
3. Press (5×) to select Ao 050.
4. Press "OK".
AnA.Mod appears on the display.
5. Press "OK".