15. Specification | 15.4 - Weighing electronics data | 186
15.4.3 Accuracy and stability
Accuracy* ≤6,000 d, acc. to OIML R76 Klasse III
Min. measuring signal * ≥0.1 mV/V = 1.2 mV corr. to 0.2 µV/d für 6,000 d
Linearity* <0,003%
Zero point stability error
(TK
0
)*
<0.05µV/K RTI ≤0.004 %/10 K @ 1 mV/V
SPAN stability error
(TK
span
)*
<±2.5ppm/K
* at a measurement time of 160 ms.
15.4.4 Sensitivity
Sensitivity 0.2 µV/d @ 6,000 d (Kl. III) OIML R76
Max. resolution 7.5 million internal counts at 3mV/V, not verifi
able
Measurement input
(measuring signal +
dead load)
U
DC
= 0…max. 36mV, symmetrical to zero
Dead load suppres
sion
max. measuring signal of U
DC
= 36mV (dead load
+ range); entry/calibration via software
15.4.5 Connecting cables
Connection PR no./cable type Length
between junction box
and device
PR 6135, PR 6135A max. 500m – length
of the load cell cable