4. Device installation | 4.3 - Hardware construction | 69
Technical data
Description Data
External connection
D‑Sub
9-pin plug connector (female)
Transfer rate 9.6kbit/s to 12Mbit/s, baud rate auto-detection
Connection mode ProfiBus network, connections can be made/re
leased without affecting other stations.
Protocol
PROFIBUS-DP-V0 SLAVE to IEC 61158
Mono or multi-master systems are supported.
Master and slave devices, max. 126 nodes possi
ble.
Watchdog Timer
Configuration GSD file "53100F3E.gsd"
Bus termination Panel device: The bus termination in the last device
is implemented via the integrated terminating resis
tor in the ProfiBus plug connector.
Table-top device: The bus termination in the last de
vice is implemented via the integrated terminating
resistor in the ProfiBus adapter board, see Chapter
ProfiBus adapter board for table-top devices.
Potential isolation Yes, optocoupler in lines A and B (RS-485)
Cable type ProfiBus "special"; color: violet; screened twisted
pair cable
Cable impedance 150Ω
Cable length The max. distance of 200m can be extended at
1.5Mbit/s by means of an additional repeater.
Certificates Profibus test center Comdec in Germany and PNO
(Profibus User Organization).
Industry-compatible CE, UL, and cUL
Note: The GSD file is available to download online:
http://www.minebea-intec.com