72
Procedures
1
Transmit the calibration start command from
the host.
Here, set the maximum load that can be applied
within the maximum value on the load cell.
The display in the load display section displays
“−RS−” and “The present load value”
alternately.
246
@ 0 0 0 5 Terminator
Command Host→CSD−891B
2 4 6 8 10 12 14
@ 0 0 0 5 S P A N Termi
ID No.
Return CSD−891B→Host
ID No.
-nator
2
Transmit the present load data reading out
command from the host.
As for the load value read by the load data
reading out command, the decimal point
adheres to the decimal point position in
F01 when you set to add the decimal point
in F−56.
246
@ 0 0 2 1 Terminator
Command Host→CSD−891B
2 4 6 8 10 12 14
@ 0 0 2 1 + 2 0 0 0 Termi
ID No.
Return CSD−891B→Host
Sign(+/−)
Present load value
ID No.
-nator
3
Transmit the display value to want to display
now by span fine adjustment command.
According to the return data, proceed to the
following steps.
・When the return data is as same as command.
As the termination is in normal, proceed to the
next step.
・When the return data is “SP−L”
The span fine adjustment range is less than 0.2
mV/V. Set the span fine adjustment range from
0.2 mV/V to 3.1 mV/V.
・When the return data is “SP−H”
The span fine adjustment range exceeds over
3.1 mV/V. Set the span fine adjustment range
from 0.2 mV/V to 3.1 mV/V.
2 4 6 8 10 12 14
@ 0 0 1 8 2 0 0 0 Termi
ID No.
Command Host
→
CSD−891B
2 4 6 8 10 12 14
@ 0 0 1 8 2 0 0 0 Termi
ID No.
Return CSD−891B→Host
・Normal termination
2 4 6 8 10 12 14
@ 0 0 1 8 S P - L Termi
ID No.
・When the span fine adjustment range is less
than 0.2 mV/V
Sign(+/−) Setting load value
Sign(+/−)
2 4 6 8 10 12 14
@ 0 0 1 8 S P - H Termi
ID No.
・When the span fine adjustment range exceeds
over 3.1 mV/V
-nator
Setting load value
-nator
-nator
-nator