120
Data Processor
Functions
Optional Setting
Input ranges for the set values
(Tolerance) Y: 0.01 to 99.99, xy: 0.0001 to 0.9999, other than Yxy: 0.05 to 80.00
(Offset) -5.00 to +5.00
(Rotation) -45 to +45
(Warning level) 10 to 100%
6
Press the Measure/Enter key
.
• The warning level setting screen appears.
7
Set the warning level using the key and the nu-
meric pad.
8
Press the Measure/Enter key
.
• Once the box tolerance setting is complete, the measurement
screen appears.
• Measurement points within the color difference tolerance are
plotted as O, warning level points are plotted as ∆, and points
outs of the color difference tolerance are plotted as X.
[TOLERANCE]
◎WARNING LEVEL
100%
P00
C
P
[0001]
T02 [0001]
ΔL* Δb*
0.5 0.5
P00
C
P
[0001]
5
Set the color difference tolerance using the key and the numeric pad.
Setting screen
Measurement result
+ input value
of ∆L*
Target
value
− input value
of ∆a*
− input value
of ∆L*
+ input value of ∆b*
+ input value of ∆a*
− input value of ∆b*