147
Description
ER17
INCORRECT
CLOCK OPERATION
Clock IC is not working correctly.
Clock IC power voltage drop or breakdown
of circuits relating to the clock IC.
Insert batteries again.
If the problem persists, contact your nearest
Minolta-authorized service facility.
ER27
ERROR
IN CHARGING
Failed to supply power to the flash circuit.
The flash circuit is broken.
Power voltage drop
Data has been deleted because the memory
backup batteries are low.
Recharge the backup battery for the memory by
turning ON the power switch. A full charge is
achieved in 20 hours and will last for approxi-
mately 10 months.
DATA SET ERROR
The data entered is outside the data set range.
Color space and data name have been entered
which is not within the range that can be pro-
cessed by the instrument.
Incorrect date entered.
overrun data.
NO TARGET COLOR
No target color has been set. Repeat after setting a target color.
USE ANOTHER
COLOR SPACE
The function being attempted does not work
with the current color space.
Statistical operations with Munsell or user in-
dexes
Target color operations when using Munsell
Manual input of target values when using
L*C*h, LCh99, WI/Tw, WI, YI, CMC,
CIE1994, or CIE2000.
Re-attempt after changing to a different color
space.
NO PAPER
The printer has run out of paper. Insert the paper.
Turn the printer setting OFF.
DATA PROTECT
IS VALID
Since the data protection setting is ON, the func-
tion being attempted is restricted.
Page deletion is impossible.
ER30
LOW BACK UP
BATTERY
VOLTAGE
Turn OFF the data protection setting before
deleting the page.
Oldest measurement data is overwritten if the
data protection setting is turned OFF.
OK02
LOW
ILLUMINATION
The xenon arc lamp has dropped to below 65%
of new illumination.
The xenon arc lamp has gotten old.
The diffusion chamber has gotten dirty.
OK05
DIFF.
ILLUMINATIONS
CAN’T
COMMUNICATE
The illumination settings for the measuring head
and the data processor are different.
Initialize them and change the illumination set-
tings.
If the problem persists, contact your nearest
Minolta-authorized service facility.
Message Symptom/Possible Cause Corrective Acrion
When using the batteries, turn OFF the power
and replace new ones.
ER12
SAMPLE TOO DARK
Low reflectivity of measurement specimen
means there is not enough light coming off the
specimen for measurement.
Redo the white calibration and the measure-
ment.
If the problem persists, contact your nearest
Minolta-authorized service facility.
ER13
ERROR IN A/D
A/D conversion failed.
The A/D converter is broken.
Circuit the peripherals is broken.
Repeat measurement.
If the problem persists, contact your nearest
Minolta-authorized service facility.