8 | en
8
Guidance and Manufacturer’s Declaration – Electromagnetic emissions
The Propulse is intended for use in the electromagnetic environment specified below. The customer
or user of the Propulse should assure that it is used in such an environment.
Emission tests Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The Propulse uses only energy for its internal function.
Therefore, its RF emissions are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A The Propulse is suitable for use in all establishments
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic
emissions IEC
61000-3-2
Class A
Voltage
fluctuation/flicker
emissions IEC
61000-3-3
Complies
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Propulse is intended for use in the electromagnetic environment specified below. The customer
or the user of the Propulse should assure that it is used in such an environment
Immunity test IEC 60601 test level Compliance level Electromagnetic environment -
guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/ burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for
input/output lines
2 kV for power
supply lines
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV differential
mode
2 kV common
mode
1 kV differential
mode
2 kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on
power supply input
lines
IEC 61000-4-11
< 5 % UT (> 95 %
dip in UT) for 0.5
cycle 40 % UT (60
% dip in UT) for 5
cycles 70 % UT (30
% dip in UT) for 25
cycles
< 5 % UT (> 95 %
dip in UT) for 5 s
< 5 % UT (> 95 %
dip in UT) for 0.5
cycle 40 % UT (60
% dip in UT) for 5
cycles 70 % UT
(30 % dip in UT)
for 25 cycles
< 5 % UT (> 95 %
dip in UT) for 5 s
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Propulse requires
continued operation during power
mains interruptions, it is
recommended that the Propulse
be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m Not applicable Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE UT is the A.C. mains voltage prior to application of the test level.