Connecting to External Devices
20
Data output timing on startup
Starting of data output
Power ON operation (parameter 17): 1, 2 5000ms max
CE key operation (parameter 17): 0 2000ms max
LF
50ms±5ms
Data will be output with 50ms interval
regardless of the transmission speed.
X +1234.567 CR LF
Data output
Next command
Last bit of command
Data
Zero set
Error cancel
100ms or less 100ms or more
Next TSP
TSP input
Data
100ms or less 100ms or more
1000ms or more
100ms or more
Min. 100ms
H level: 4.0 - 5.0V
L level: 0.0 - 0.5V
TIMING CHART
4RIGGER¬-ODE¬+!¬COUNTER¬AND¬+,$ ¬COUNTER
Measurement data can be output by commands from the
computer.
)NTERVAL¬-ODE¬+!¬COUNTER
Measurement data can be output at specific intervals.
The display mode at startup can be selected with
THE¬PARAMETER¬¬+!¬COUNTER
When requesting data repeatedly from a touch-
SIGNAL¬PROBE¬ALLOW¬¬SECOND¬OR¬MORE¬BETWEEN¬INPUT¬
signals.
4RIGGER¬-ODE¬+!¬COUNTER¬AND¬+,$ ¬COUNTER
Measurement data can be output by the signal from a touch-signal
probe and external load box.
%XTERNAL¬ZERO¬SIGNAL¬+!¬COUNTER¬AND¬+,$ ¬COUNTER
-AINTAIN¬THE¬LOW¬LEVEL¬FOR¬MS¬OR¬MORE
Each value in the timing chart indicates response
time to a command. Consequently, be aware that
this command may cause a difference between a
detected point and the actual one when the slider is
moving.
"ECAUSE¬THE¬+!¬COUNTER¬DOES¬NOT¬USE¬A¬LATCH¬AT¬
count sampling under conditions of normal data
output (by command or touch-signal probe), an
inaccurate data output may be produced when a
data request is made while moving the slider relative
to the scale at 70mm/s or more.