GeneralOperation
Stage Preparation
1.StageInstallation.
1.1. RemoveaSeebeckStagefromthepackage.
1.2. PutathinlayerofthermalgreaseonthelargeCopperplate(underside).
1.3. Unlock the ZIF connector (if present) on the preamplifier board in the vacuum chamber. Using
thumbsoneachsideoftheconnecter,
inserttheSeebeckstagefully.
1.4. Affix the Seebeck Stage to The Cold Stage by the springfasteners. Be extremely careful! Do not
applyanydownwardforceontherefrigerator.Usetweezerstoadjustthefastenersduringinstallation.
You can move the Seebeck Stage backwards and forwards by a couple
of millimeters to get the best
location.
1.5. LocktheZIFconnector(ifpresent).
1.6. Installthecoveronthevacuumchamber.
2. ConnecttheRibboncablefromtheChambertotheSB1000.IfaK2000ispresentfortemperature
control,ensuretheLinkRibbonCableisinstalledfromthe
SB1000totheK2000.
3. If you are going to use the system at temperatures below ambient, you should also install all
systemcomponentsrequiredforcoolingaccordingtotheMMRRefrigeratorsManual.