ImageXpress Confocal HT.ai High-Content Imaging System User Guide
12 5084773 A
Electrical Safety
To prevent electrically related injuries and property damage, inspect all electrical equipment
before use and immediately report all electrical deficiencies. Contact Molecular Devices
Technical Support to service equipment that requires the removal of covers or panels.
We recommend that you power off the instrument when it is not in use.
WARNING!
The ImageXpress Confocal HT.ai system is an Equipment Class 1 product that
relies on protective earth grounding for safe operation. Any interruption of the
protective earth ground conductor—inside or outside the instrument—or
disconnection of the protective earth ground terminal can result in personal injury.
Do not position the equipment so that it is difficult to operate the power switch on
the front of the ImageXpressSystems Power and Options Controller.
WARNING! HIGH VOLTAGE. Do not operate the external light source with the
external light source housing open. Never attempt to open or tamper with the laser
light source housing due to the extreme hazard of the Class 4 high-power laser
product.
ImageXpressSystems Power and Options Controller
A single power cable connects the instrument to the external ImageXpressSystems Power
and Options Controller. The external power controller has an input voltage rating of 100VACto
240VAC, 50/60Hz, 12amps maximum. The power controller contains no user-serviceable
parts.
WARNING! Before attempting to access any internal service areas of the instrument ,
unplug the power cable.
Fuses and Circuit Protection
In the ImageXpress Confocal HT.ai system, the ImageXpressSystems Power and Options
Controller provides over-current protection for the light source limited to 15amps maximum.
The power controller contains no user-serviceable parts.