IPA287Supplement Page 6 Monroe Electronics, Inc.
Error Messages What to do...
Battery Low 6.2
V
Off in 5 sec.
---V A30%M50%
28C 9.0V Er%RH
Change the battery.
Humidity is out of 10% - 90% range or
sensor failure. If sensor failure, contact
factory for repair.
+1012V +Charge
Unable to charge
Unable to charge to 1050V. Make sure nothing is contacting the charge plate
and the standoffs underneath are clean. This may also be caused by too
much load on the plate. Press enter to return to the main screen. Remove
the 287 from the ionizer and recheck. If problem persists contact the
manufacturer for calibration.
+ 216V 200.s
Decay failed
Decay time out at 200 seconds, and the plate reading is still above
100V. Check your ionizer. Recommend using Monroe Model 288
EEProm Fail
Contact factory.