Specifications
Range: 1.0 x 10
3
to 1 x 10
12
Ω @ 10V
1.0 x 10
7
to 1 x 10
12
Ω @ 100V
Accuracy: 1.0 x 10
3
to 1 x 10
5
Ω @ 10V ±10% of reading at 70-80
0
F; 15–60% RH
1.0 x 10
7
to 1 x 10
12
Ω @ 100V ±10% of reading at 70-80
0
F; 15–60% RH
Open Circuit Voltages: 10 and 100 volts ± 3%
Electrification Period: 15 seconds
Temperature Accuracy: ± 5 degrees F and ± 3 degrees C Typ.
Relative Humidity: Range from 5% - 95%; ± 10% Typ.
Power: Two - Alkaline AA batteries
Battery Life: approximately 1500 measurements.
Meter Jacks: Left jack on meter accepts a 3.5 mm plug
Right jack on meter accepts a standard banana plug
Test Leads: Black lead terminated with a 3.5 mm plug one end and a standard banana
plug on the other
White lead terminated with a standard banana plus both ends
External Electrodes: Two NFPA-99 5-pound electrodes ( 80 oz.) ± 2oz with Shore A (IRHD)
50 – 70 durometer conductive pads that comply with EOS/ESD S4.1
Resistivity Electrodes: Parallel conductive silicone rubber electrodes 4 inches long and spaced
4 inches apart mounted onto an adapter sled
Meter Weight: 15.5 oz. With out adapter; 23 oz. with adapter
Dimensions: 8” L x 4.3” W x 1.6” H
Kit includes:
! Meter
! Protective carrying case
! Test leads
! Two NFPA-99 5-pound electrode weights
! Two AA Alkaline batteries
! Operator’s Manual
Resistivity adapter sled is optional and must be ordered separately.
Specifications are subject to change without notice.