Chapter 3
Specifications
Table 3-1. Performance Specifications
GAS TYPES
Combustibles, Oxygen & Toxics
TEMPERATURE
RANGE
TOXICS &
OXYGEN
OPERATING
RANGE
0 to 40
o
C
(32 to +104
o
F)
*EXTENDED
RANGE
-20 to +50
o
C
(-4 to +122
o
F)
OPERATING
RANGE NH
3
0 to +30
o
C
(32 to +86
o
F)
*EXTENDED
RANGE NH
3
, CL
2
,
CLO
2
-10 to +40
o
C
(-14 to +104
o
F)
Calibrate within operating range
CATALYTIC
COMBUSTIBLES
SINGLE MODULE
-40 to +60
o
C
(-40 to +140
o
F)
DUAL MODULE
-40 to +60
o
C
(-40 to +140
o
F)
IR
COMBUSTIBLES
SINGLE MODULE
-40 to +60
o
C
(-40 to +140
o
F)
DUAL MODULE
DRIFT
ZERO DRIFT
Less than 5%/year, typically
SPAN DRIFT
Less than 10%/year, typically
NOISE
Less than 1% FS