Model 493.02 Controller Service
Calibrating an LVDT
Calibration
122
7. On the Linearization Data window, enter the dial indicator and
conditioner feedback readings for all command points previously
recorded.
Stand-Alone— Path: Setup > Disp > Calibration > Cal Type >
Gain/Linearization > Linearization Data
Automated— Click Linearization Data on the Calibration tab
to open the Linearization Data window.
8. Turn on system hydraulics.