Part 12470
Rev. 2 3/25/99
Initiates the output circuit in conjunction with the appropriate Timer
STOP MODE Switch position.
Momentary on position with spring return to OFF. Output will
remain energized as long as switch is held in MOMENTARY
position and is de-energized when released. Timer starts and stops in
the same manner. Used for jogging the output circuit.
Maintained on position with spring return to OFF. Output will
remain energized until appropriate STOP MODE function occurs,
switch is moved to OFF position or current drops below threshold
requirement.
Test set input power remains energized but output current remains
de-energized and timer will not count.
LED displays show the numeral one in the left most position when
the meter is over-ranged.
Decimal of display shifts one digit to the right for each increase in
range selection.
Selects either the NORMAL or MEMORY mode of the ammeter.
Ammeter updates reading continually as long as the output is
energized. Reading is lost when the output is de-energized.
Ammeter retains the highest peak value attained during current
output operation above 8% of the full scale of range selected.
Reading is retained until the output is reinitiated or RANGE
SWITCH position is changed.
Automatically resets each time output is reinitiated.
Selects cycles or seconds mode.