IBIC Section Five
GPIB-PC User Manual 5-26 ©National Instruments Corp.
Example:
GPIB0:
[01F4] (cmpl lok rem cic atn lacs)
count: 1
ibcmd "\x19"
Since the DVM and the GPIB-PC are still addressed to talk and to
listen, the measurement can be read.
Example:
GPIB0:
[01E4] (cmpl lok rem cic lacs)
0D 0A 4E 44 43 56 2D 30• • N D C V - 0
30 30 2E 30 30 34 37 450 0 . 0 0 4 7 E
2B 30 0D 0A + 0 • •
ibrd 20
To close out a programming sequence, send the interface clear message
(IFC) to initialize the bus.
Example:
GPIB:
[0160} (cmpl lok rem cic)
ibsic
Return to DOS.