TROUBLESHOOTING
Issue Action
Motor isn’t running or
motor is running slow
Ensure batteries are installed correctly.
Replace batteries.
I don’t feel any difference 1. Ensure you have an airtight fit between the nosepiece
and your nostril.
2. Ensure you have your other nostril pinched tightly.
3. Ensure you are swallowing while the Eustachi is running.
4. If still no difference, verify there is no air leak (see below).
5. If still no difference, contact your doctor
Suspected air leak 1. Push the power button and at the same time
plug the nosepiece hole with your finger.
2. The sound of the motor should change when
you plug the hole. If you don’t hear a change
in the motor, there may be an air leak.
ELECTRICAL SAFETY INFORMATION
Electrical classification:
Internally powered ME Equipment
Ingress protection rating: IP22. The first digit, 2,
indicates protection against a solid object greater
than 12.5mm such as a finger. The second digit, 2,
indicates protection against vertical falling drops
of water with enclosure tilted up to 15 degrees.
Duty cycle: 30 seconds on / 2 minutes off
Operating Environment:
Temperature: 41°F to 104°F (5°C to 40°C)
Humidity: 15 to 90%
Atmospheric Pressure: 21 to 32 inHg
(700 hPa to 1,060 hPa)
Storage and Transportation Environment:
Temperature: -13°F to 158°F (-25°C to 70°C)
Humidity: up to 90%
Intended Operator: Adult patient or adult caregiver
Applied Part: air outlet (hole) of the nosepiece
SYMBOLS GLOSSARY
Symbol Std Reference Definition
ISO 15223-1: 2016 symbol 5.4.4 Caution see accompanying documents
ISO 15223-1: 2016 symbol 5.4.3 Consult instructions for use
ISO 15223-1: 2016 symbol 5.1.6 Catalog, model, part number
ISO 15223-1: 2016 symbol 5.1.5 Lot (batch) number
ISO 15223-1: 2016 symbol 5.1.3 Date of manufacture (YYYY-MM-DD)
ISO 15223-1: 2016 symbol 5.1.1 Manufacturer
ISO 7000: 2019 symbol 2725 Not made with natural rubber latex
ISO 15223-1: 2016 symbol 5.2.7 Non-sterile
ISO 15223-1: 2016 symbol 5.3.4 Keep dry
IEC 60529 Ingress protection rating. The first digit, 2, indicates protection against a solid object greater
than 12.5mm such as a finger. The second digit, 2, indicates protection against vertical falling
drops of water with enclosure tilted up to 15 degrees.
IEC 60601-1 Table D.1 Symbol 20 Type BF applied part: electrically connected to patient but not directly connected to the heart
ISO 15223-1: 2016 symbol 5.3.7 Temperature limits
ISO 15223-1: 2016 symbol 5.3.8 Humidity limits
ISO 15223-1: 2016 symbol 5.3.9 Atmospheric pressure limits
IEC 60601-1 Table D.1 symbol 29 Stand-by; press and hold to power on device
EN 50419 Do not throw in trash
5003FAN-EUS-US-Rev5 Page 2 of 2
EXPECTED PRODUCT LIFE
Your Eustachi should last for up to 5 years with proper care.
WARRANTY
This Eustachi is warranted to be free from manufacturing defects for a period of one year from the date of purchase under normal use. This excludes damage to the product
resulting from accident or misuse. CONTACT INFORMATION
For assistance in setting up, using or maintaining the Eustachi, to report unexpected operation or events, or for other questions or comments, please go to our website:
www.eustachi.com or email us at info@eustachi.com