NeuMoDx™ HCV External Controls
INSTRUCTIONS FOR USE
900201
NeuMoDx Molecular, Inc. CONFIDENTIAL P/N 40600196_Rev B
March 2019
Pg. 2 of 4 For Ex-US Distribution Only
NeuMoDx™ Reagents and Consumables Required But Not Provided
Other Equipment and Materials Required But Not Provided
NeuMoDx 288 Molecular System [REF 500100] or NeuMoDx™ 96 Molecular System [REF 500200]
WARNINGS & PRECAUTIONS
The NeuMoDx HCV External Controls are for in vitro diagnostic use only with the NeuMoDx HCV Quant Test Strip as implemented on
the NeuMoDx System.
Do not use the NeuMoDx HCV External Controls after the listed expiration date.
Do not use the NeuMoDx HCV External Controls if the packaging is damaged or the kit is not frozen upon arrival.
Because the NeuMoDx HCV positive controls contain intact HCV positive material, all components of the NeuMoDx HCV External
Controls should be treated as potentially biohazardous.
Always handle specimens as if they are infectious and in accordance with safe laboratory procedures such as those described in
Biosafety in Microbiological and Biomedical Laboratories
1
and in CLSI Document M29-A3.
2
Do not pipette by mouth. Do not smoke, drink, or eat in areas where specimens or reagents are being handled.
Dispose of unused reagents and waste in accordance with country, federal, provincial, state and local regulations.
Clean, powder-free, nitrile gloves should be worn when handling all NeuMoDx reagents and consumables.
Wash hands thoroughly after performing the test.
Safety Data Sheets (SDS) are provided for each reagent.
PRODUCT STORAGE, HANDLING & STABILITY
The NeuMoDx HCV External Controls are shipped with dry ice to maintain a frozen state; do not use if kit contents are not frozen upon
receipt.
It is recommended that the NeuMoDx HCV External Controls be stored at -80°C to ensure stability.
Control vials are intended for single use only. Thawed external controls may be stored at 4°C for no longer than 24 hours.
Refreezing after a first thaw is not recommended.
Discard any unused material after use as biohazard waste.
Discard any controls that appear cloudy or contain large precipitates after thawing.
REF
Contents
300300
NeuMoDx™ HCV Quant Test Strip
Dried PCR reagents containing HCV-specific TaqMan
®
probes and primers along with SPC2-specific TaqMan® probe and primers.
100200
NeuMoDx™ Extraction Plate
Dried magnetic affinity microspheres, Lytic enzyme, and sample process controls
800200
NeuMoDx™ HCV Calibrators
Single use sets of HCV High and Low Calibrators to establish validity of standard curve
400600
NeuMoDx™ Lysis Buffer 3
400100
NeuMoDx™ Wash Reagent
400200
NeuMoDx™ Release Reagent
100100
NeuMoDxCartridge
235903
Hamilton CO-RE Tips (300 µL) with Filters (available from NeuMoDx or Hamilton)
235905
Hamilton CO-RE Tips (1000 µL) with Filters (available from NeuMoDx or Hamilton)