25
GPIB Bus SW SW SW SW SW
Address 1 2 3 4 5
1: 10000
2: 01000
3: 11000
4: 00100
5: 10100
6: 01100
7: 11100
8: 00010
9: 10010
10: 0 1 0 1 0
11: 1 1 0 1 0
12: 0 0 1 1 0
13: 1 0 1 1 0
14: 0 1 1 1 0
15: 1 1 1 1 0
16: 0 0 0 0 1
17: 1 0 0 0 1
18: 0 1 0 0 1
19: 1 1 0 0 1
20: 0 0 1 0 1
21: 1 0 1 0 1
22: 0 1 1 0 1
23: 1 1 1 0 1
24: 0 0 0 1 1
25: 1 0 0 1 1
26: 0 1 0 1 1
27: 1 1 0 1 1
28: 0 0 1 1 1
29: 1 0 1 1 1
30: 0 1 1 1 1
Default GPIB Address=4
1 2 3 4 5 6 7 8
1
0
GPIB Address
RS232C
Not
Used
Figure 9. GPIB Address Selector Switches
5.5.2 GPIB Command Termination:
Commands and queries sent to the 1830-C through the GPIB bus should be
terminated by sending an <NL><EOI> (<NL> is equivalent to an <LF>).
All responses sent by the 1830-C are terminated with the concurrent
transmission of a <NL><EOI>.