6.2 Detailed Description of Commands and Queries
An
Attenuator on/off
Function: Selects between the responsitivity values associated with
the photodetector alone or for the photodetector-attenuator
combination. These values are stored within the calibration
module.
Syntax: An or an
Parameter: n = 0 Use the calibration module’s responsitivity values
associated with the photodetector-alone.
n = 1 Use the calibration module’s responsitivity values
associated with the photodetector-attenuator
combination.
Type: integer
Related
Commands: A? – Attenuator query
A?
Attenuator query
Function: Reports whether the 1830-C is using the responsitivity
associated with the photodetector-alone or the
photodetector-attenuator combination.
Syntax: A? or a?
Returns: 0 photodetector-alone responsitivities are used.
1 photodetector-attenuator responsitivities are used.
Related
Commands: An – Selects the responsitivity associated with the
photodetector-alone or photodetector-attenuator combination