Configuring The Meter
4.7 SETPOINT 2 DEADBAND (S2.DB)
Setpoint 2 Deadband is not active unless your meter has dual relay output capabilities. The
LEDÕs will display whether the (S2.DB) is active or not. To set the deadband (hysteresis) of
Setpoint 2, follow these steps:
1. Press the MENU button until the meter shows "S2.DB".
2. Press the
©
/DEV button. The meter shows the last previously stored 4-digit number
(0000 through 9999) with flashing 4th digit.
3. Press the
¶
/MAX button to change the value of the flashing digit. If you continue to
press the
¶
/MAX button, the flashing digit's value continues to change.
4. Press the
©
/DEV button to scroll to the next digit.
5. Press the MENU button to store your selection. "STRD" momentarily displays, followed
by cold junction offset if you have a standard meter (refer to Section 4.12) or "OT.CF"
(Output Configuration) if you have analog output capabilities (refer to Section 4.8).
4.8 OUTPUT CONFIGURATION (OT.CF)
Output Configuration is not active unless your meter has analog output capabilities. The menu
will display whether analog output is present or not. Use Output Configuration (OT.CF) to select
the following:
* To enable or disable the analog output
* To determine if the analog output is current or voltage
* To determine if the analog output is a retransmission of the display or proportional to the
error (the difference between reading and setpoint value)
4.8.1 To Enable or Disable The Analog Output
1. Press the MENU button until the meter shows "OT.CF" .
2. Press the
©
/DEV button. The meter shows one of the following:
* "O.1=D" (Analog output disabled)
* "O.1=E" (Analog output enabled)
3. Press the
¶
/MAX button to toggle between choices.
4. Press the
©
/DEV button to select analog output as current/voltage or press the MENU
button to store your choice. The meter momentarily shows "STRD", followed by "OT.S.O"
(Output Scale and Offset - refer to Section 4.11).
22