11. Press the
©
/DEV button. The last previously stored 4-digit number (-1999 through 9999)
displays with flashing 4th digit.
12. Press the
¶
/MAX button to change the value of the flashing digit. If you continue to
press the
¶
/MAX button, the flashing digit's value continues to change.
13. Press the
©
/DEV button to scroll to the next digit.
14. Press the MENU button to store your selection. "OUT.2" (Output 2) displays.
This analog signal should correspond to your Read 2 display.
15. Press the
©
/DEV button. Selected output displays.
If you select "O.2=V" for voltage, the maximum signal you may select is 10.00 for an
0 to 10 V dc signal output. If you select "O.2=C" for current, the maximum signal you
may select is 19.99 for a 0 to 20 or 4 to 20 mA dc signal output.
16. Press the
¶
/MAX button to change the value of the flashing digit. If you continue to
press the
¶
/MAX button, the flashing digit's value continues to change.
17. Press the
©
/DEV button to scroll to the next digit.
18. Press the MENU button to store your selection. The meter momentarily shows "STRD",
followed also momentarily by "RST" (Hard Reset). Meter then returns you to the "RUN"
mode.
WARNING: If the meter displays all flashing values on any item, the value has
overflowed. Press the
¶
/MAX button to start new values.