What to do if Nidek Medical Laboratory Equipment ICE-1000 shows Optical Initialization Error?
J
jennifer88Aug 5, 2025
If you encounter an Optical Initialization Error on your Nidek Medical Laboratory Equipment, it indicates a grid detection error during power-up. First, remove any foreign substances or soiling from the lens table and restart the blocker. If the issue persists, detach the lens table and restart the blocker. Once the blocker restarts successfully, reattach the lens table.
N
Nicholas SmithAug 13, 2025
What to do if Nidek Medical Laboratory Equipment ICE-1000 does not work at all?
K
Kristen LeonAug 13, 2025
If your Nidek Medical Laboratory Equipment ICE-1000 is not working, check the following:
* Ensure the power cord is properly connected.
* Verify the outlet voltage is within the specified range.
* Inspect the fuses and replace them if they are blown. Refer to section “3.2 Replacing Fuses” (Page 86) for fuse replacement instructions.
L
Linda GrossAug 20, 2025
Why is the measured data of the lens unstable on Nidek Medical Laboratory Equipment?
B
Barbara MillerAug 20, 2025
Unstable lens measurement data on your Nidek Medical Laboratory Equipment may be due to dirt on the lens table (grid plate). Wipe the lens table clean. Also, ensure the instrument is not exposed to direct illumination.
R
Ronnie CruzAug 28, 2025
What to do if Nidek Medical Laboratory Equipment shows X Movement Error, Y Movement Error, R Movement Error, Reverse Movement Error, Block Movement Error, Clamp Error, Upper Pin Operation Error, Lower Pin Operation Error?
J
Joe AndrewsAug 28, 2025
If your Nidek Medical Laboratory Equipment ICE-1000 displays any of the following errors: X Movement Error, Y Movement Error, R Movement Error, Reverse Movement Error, Block Movement Error, Clamp Error, Upper Pin Operation Error, or Lower Pin Operation Error, it indicates an issue with the blocking mechanism. After noting the specific error message, shut down the ICE-1000 and contact NIDEK or your authorized distributor for assistance.